Abstract:
This work presents the development of an intelligent microsurgical instrument to perform real-time tremor compensation within a handheld tool. The intelligent instrument senses its own motion, distinguishes between voluntary and erroneous motion, and manipulates its tip to cancel the undesired component in real-time. The on-board sensing unit is made up of a magnetometer-aided all-accelerometer inertial measurement unit and sensor fusion is performed via a quaternion-based Kalman filtering. Tremor is modeled and filtered by an adaptive zero-phase notch filter. The intraocular shaft manipulator is a three DOF piezoelectric actuated mechanism driven by a feedforward controller with inverse rate-dependent hysteresis model. Laboratory experimental results of the system are presented.
copied = false, 2000);
">
@conference{Ang-2004-16951,
author = {Wei-Tech Ang And Pradeep Khosla And Cameron Riviere},
title = {Active tremor compensation in microsurgery},
booktitle = {Proceedings of 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC '04)},
year = {2004},
month = {September},
pages = {2738 - 2741},
}
author = {Wei-Tech Ang And Pradeep Khosla And Cameron Riviere},
title = {Active tremor compensation in microsurgery},
booktitle = {Proceedings of 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC '04)},
year = {2004},
month = {September},
pages = {2738 - 2741},
}