Factorization-Based Calibration Method for MEMS Inertial Measurement Unit

May 2008

Factorization-Based Calibration Method for MEMS Inertial Measurement Unit

Conference Paper
Proceedings of (ICRA) International Conference on Robotics and Automation

Abstract:

We present an easy-to-use calibration method for MEMS inertial sensor units based on the Factorization method which was originally invented for shape-and-motion recovery in computer vision. Our method requires no explicit knowledge of individual motions applied during calibration procedure. Instead a set of motion constraints in the form of an innerproduct is used to factorize sensor measurements into a calibration matrix (that represents intrinsic sensor parameters) and a motion matrix (that represents acceleration or angular velocity). These motion constraints can be collected quickly from a lowcost calibration apparatus. Our method is not limited to just triad configurations but also applicable to any coordination of more than three sensor elements. A redundant configuration has the benefit that all the calibration parameters including biases are estimated at once. Simulation and experiments are provided to verify the proposed method.
@conference{Hwangbo-2008-9972,
author = {Myung Hwangbo And Takeo Kanade},
title = {Factorization-Based Calibration Method for MEMS Inertial Measurement Unit},
booktitle = {Proceedings of (ICRA) International Conference on Robotics and Automation},
year = {2008},
month = {May},
pages = {1306 - 1311},
keywords = {IMU calibration, Factorization method},
}
Copyright notice: This material is presented to ensure timely dissemination of scholarly and technical work. Copyright and all rights therein are retained by authors or by other copyright holders. All persons copying this information are expected to adhere to the terms and constraints invoked by each author's copyright. These works may not be reposted without the explicit permission of the copyright holder.