Microscale Tracking of Surgical Instrument Motion

September 1999

Microscale Tracking of Surgical Instrument Motion

Authors:

Cameron Riviere and Pradeep Khosla

Abstract:

An apparatus for accurate three-dimensional tracking of the tip of a microsurgical instrument has been developed for laboratory use. The system is useful for evaluation of microsurgical instrument designs and devices for accuracy enhancement (both robotic devices and active hand-held instruments), as well as for assessment and training of microsurgeons. It can also be used as a high-precision input interface to microsurgical simulators. The system involves illumination of the workspace and optical sensing of the position of a small reflective ball at the instrument tip, and therefore requires no wiring connection to the instrument being tracked. Sensing is performed by two position-sensitive photodiodes, placed orthogonally. The rms noise per coordinate is presently 7 microns. Preliminary results are presented. The photodiodes exhibit some degree of nonlinearity, which can be calibrated. The goal is to achieve an rms noise level of 1 micron. This is expected to be attainable via a synchronous detection scheme which has not yet been implemented.
@conference{Riviere-1999-15016,
author = {Cameron Riviere And Pradeep Khosla},
title = {Microscale Tracking of Surgical Instrument Motion},
booktitle = {Proceedings of 2nd International Conference on Medical Image Computing and Computer-Assisted Intervention (MICCAI '99)},
year = {1999},
month = {September},
pages = {1080 - 1087},
}
Copyright notice: This material is presented to ensure timely dissemination of scholarly and technical work. Copyright and all rights therein are retained by authors or by other copyright holders. All persons copying this information are expected to adhere to the terms and constraints invoked by each author's copyright. These works may not be reposted without the explicit permission of the copyright holder.